HI83314 is a compact, multiparameter photometer for measuring key water and wastewater quality parameters. The meter is one of the most advanced photometers available with an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. 10 key water and wastewater quality parameters with 20 different methods covering multiple ranges are programmed into the meter. Wastewater treatment digestion parameters include COD, Total Nitrogen and Total Phosphorous which are important for monitoring nutrient removal. The HI83314 also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves.

To save valuable laboratory benchtop space, the HI83314 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.

Advanced optical system

Unparalleled performance from a benchtop photometer

Digital pH electrode input

Save valuable bench space with one meter that works both as a photometer and as a laboratory pH meter

Wastewater Treatment Digestion Parameters

Measures common wastewater parameters including COD, Total Nitrogen and Total Phosphorus

view compatible reagents

4,324 AED

Estimated delivery: 10 - 11 September

SKU:HI83314-02

HI83314 benchtop photometer measures 10 different key wastewater quality parameters using 20 different methods that allow for multiple ranges and variations in chemistry for specific applications. The Chemical Oxygen Demand (COD) parameter is included for industrial and municipal wastewater treatment. The Phosphorous and Nitrogen parameters included are beneficial to municipal wastewater treatment customers that need to monitor their biological and chemical nutrient removal process. This photometer features an innovative optical system that uses LED’s, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.

A digital pH electrode input is provided allowing the user to measure pH by a traditional glass electrode. The digital pH electrode has a built in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.

Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.

The HI83314 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the 4 wavelengths of light (420 nm, 466 nm, 525 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for education to teach the concept of absorbance by using the Beer-Lambert Law.

Features/Benefits:

Backlit 128 x 64 Pixel Graphic LCD Display

  • Backlit graphic display allows for easy viewing in low light conditions
  • The 128 x 64 Pixel LCD allows for a simplified user interface with virtual keys Ā and on-screen help to guide the user through use of the meter

Built-in Reaction Timer for Photometric Measurements

  • The measurement is taken after the countdown timer expires.
  • Countdown timer ensures that all readings are taken at the appropriate reaction intervals regardless of user for better consistency in measurements

Absorbance mode

  • Hanna’s exclusive CAL Check cuvettes for validation of light source and detector
  • Allows for the user to plot concentration versus absorbance for a specific wavelength for use with user supplied chemistry or for teaching principles of photometry

Units of Measure

  • Appropriate unit of measure along with chemical form is displayed along with reading

Result Conversion

  • Automatically convert readings to other chemical forms with the touch of a button

Cuvette Cover

  • Aids in preventing stray light from affecting measurements

Digital pH Electrode Input

  • Measure pH and temperature with a single probe
  • Good Laboratory Practice (GLP) to track calibration information including date, time, buffers used, offset and slope for traceability
  • pH CAL Check alerts user to potential problems during the calibration process
  • Space saving having a pH meter and photometer built into one meter

Data Logging

  • Up to 1000 photometric and pH readings can be stored by simply pressing the dedicated LOG button. Logged readings are just as easily recalled by pressing the RCL button
  • Sample ID and User ID information can be added to a logged reading using alphanumeric keypad

Connectivity

  • Logged readings can be quickly and easily transferred to a flash drive using the USB-A host port or to a computer using the micro USB-B port
  • Data is exported as a .CSV file for use with common spreadsheet programs

Battery Status Indicator

  • Indicates the amount of battery life left

Error Messages

  • Photometric error messages include no cap, high zero, and standard too low
  • pH calibration messages include clean electrode, check buffer and check probe

On-screen Features

method selection

Method Selection
Users can easily select any one of the 20 measurement methods and 4 absorbance methods via the dedicated METHOD button.

data logging

Data Logging
Up to 1000 measurement readings can be logged with user and sample ID and recalled for future use.

ph measurement mode

pH Measurement Mode
Selecting the pH measurement mode allows for the photometer to be used as a professional pH meter with many features including temperature compensated measurements, automatic two point calibration, and GLP.

Digestion Parameters Features

HI83399 cuvette adapter

Cuvette Adapter
The HI83314 is supplied with a 16 mm cuvette adapter that accepts digestion vials.

COD reagents

Digestion Vial Methods
Compatible with COD (EPA, ISO, and mercury free methods), Nitrogen and Phosphorous reagetns packaged in 16 mm digestion vial. Reagents are sold separately.

HI839800 COD reactor

COD Reactor for Digestion Vials
A COD reactor is used to heat the digestion vials. The digestion vials must be heated to a specific temperature for a period time making the HI839800 an important accessory required to have a complete wastewater treatment monitoring system.

Advanced Optical System

HI83314 is designed with an innovative optical system that incorporates a beam splitter so that light can be used for absorbance readings and for a reference detector. The reference detector monitors the intensity of light and modulates when there is drift due to power fluctuation or the heating of the optical components. Each part has an important role in providing unparalleled performance from a photometer.

photometer multiparameter optical

High Efficiency LED Light Source

An LED light source offers superior performance as compared to a tungsten lamp. LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce very little heat, which could otherwise affect the optical components an electronic stability. LEDs are available in a wide array of wavelengths, whereas tungsten lamps are supposed to be white light (all wavelengths of visible light) but actually have a poor blue/violet light output.

High Quality Narrow Band Interference Filters

The narrow band interference filter not only ensure greater wavelength accuracy (Ā± 1 nm) but are extremely efficient. The filters used allow up to 95% of the light from the LED to be transmitted as compared to other filters that are only 75% efficient. The higher efficiency allows for a brighter, stronger light source. The end result is higher measurement stability and less wavelength error.

Reference Detector for a Stable Light Source

A beam splitter is used as part of the internal reference system of the HI83314 photometer. The reference detector compensates for any drift due to power fluctuations or ambient temperature changes. Now you can rely on a stable source of light between your blank (zero) measurement and sample measurement.

Large Cuvette Size

The sample cell of the HI83314 fits a round, glass cuvette with a 25 mm path length. Along with the advanced optical components, the larger size of the cuvette greatly reduces errors in rotation from the indexing mark of the cuvettes. The relatively long path length of the sample cuvette allows the light to pass through more of the sample solution, ensuring accurate measurements even in low absorbance samples.

Focusing Lens for Greater Light Yield

Adding a focusing lens to the optical path allows for the collection of all of the light that exits the cuvette and focusing the light on the silicon photo detector. This novel approach to photometric measurements cancels the errors from imperfections and scratches present in the glass cuvette eliminating the need to index the cuvette.

Specifications

SKU HI83314-02
Name Wastewater Multiparameter (with COD) Benchtop Photometer and pH meter – HI83314
pH Range Photometer: 6.5 to 8.5 pH
pH electrode: -2.00 to 16.00 pH
pH Resolution Photometer: 0.1 pH
pH electrode: 0.01 pH
pH Accuracy Photometer: Ā±0.1 pH
pH electrode: Ā±0.01 pH
pH Calibration Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
pH Temperature Compensation Automatic (-5.0 to 100.0 ℃; 23.0 to 212.0 °F); limits reduced based on the pH electrode used
pH CAL Check (electrode diagnostics) clean electrode and check buffer/check probe displayed during calibration
pH Method Photometer: phenol red
pH-mV Range Ā±1000 mV
pH-mV Resolution 0.1 mV
pH-mV Accuracy Ā±0.2 mV
Absorbance Range 0.000 to 4.000 Abs
Absorbance Resolution 0.001 Abs
Absorbance Accuracy Ā±0.003Abs @ 1.000 Abs
Ammonia Range Low Range: 0.00 to 3.00 mg/L
Low Range (16 mm vial): 0.00 to 3.00 mg/L
Medium Range: 0.00 to 10.00 mg/L
High Range: 0.0 to 100.0 mg/L
High Range (16 mm vial): 0.0 to 100.0 mg/L
(all as NH3-N)
Ammonia Resolution 0.01 mg/L; 0.1 mg/L
Ammonia Accuracy Low Range: Ā±0.04 mg/L Ā±4% of reading
Low Range (16 mm vial): Ā±0.10 mg/L or Ā±5% of reading
Medium Range: Ā±0.05 mg/L Ā±5% of reading
High range: Ā±0.5 mg/L Ā±5% of reading
High range (16 mm vial): Ā±1 mg/L or Ā±5% of reading
Ammonia Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method
Chemical Oxygen Demand Range Low Range: 0 to 150 mg/L
Medium Range: 0 to 1500 mg/L
High Range:0 to 15000 mg/L
Chemical Oxygen Demand Resolution Low and Medium Range: 1 mg/L
High Range: 10 mg/L
Chemical Oxygen Demand Accuracy Low Range: Ā±5 mg/L Ā±5% of reading
Medium Range: Ā±15 mg/L Ā±4% of reading
High Range: Ā±150 mg/L Ā±3% of reading
Chemical Oxygen Demand Method Adaptation of the USEPA 410.4
ISO dichromate methods
Mercury-free dichromate green method (LR & MR); dichromate method (HR)
Total Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Total Chlorine Resolution 0.01 mg/L
Total Chlorine Accuracy Ā±0.03 mg/L Ā±3% of reading
Chlorine Method Adaptation of the EPA 330.5 DPD method
Nitrate Range 16 mm vial: 0.0 to 30.0 mg/L (as NO3– N)
Nitrate Resolution 0.1 mg/L
Nitrate Accuracy 16 mm vial: Ā±1.0 mg/L or Ā±3% of reading
Nitrate Method Adaptation of the cadmium reduction method
Nitrite Range Freshwater
Low Range: 0 to 600 µg/L (as NO2-N)
High Range: 0 to 150 mg/L (as NO2)
Nitrite Resolution Freshwater: 1 µg/L; 1 mg/L
Nitrite Accuracy Freshwater
Low Range: Ā±20 µg/L Ā±4% of reading
High Range: Ā±4 mg/L Ā±4% of reading
Nitrite Method Adaptation of the EPA Diazotization method 354.1
Nitrogen, Total Range Low Range: 0.0 to 25.0 mg/L (as NO3-N)
High Range: 0 to 150 mg/L (as N)
Nitrogen, Total Resolution 0.1 mg/L; 1 mg/L
Nitrogen, Total Accuracy Low Range: Ā±1.0 mg/L or Ā±5% of reading
High Range: Ā±3 mg/L or Ā±4% of reading
Nitrogen, Total Method Chromotropic acid method
Phosphorous, Acid Hydrolyzable Range 0.00 to 1.60 mg/L (as P)
Phosphorous, Acid Hydrolyzable Resolution 0.01 mg/L
Phosphorous, Acid Hydrolyzable Accuracy Ā±0.05 mg/L or Ā±5% of reading
Phosphorous, Acid Hydrolyzable Method Adaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 4500-P E, ascorbic acid method
Phosphorous, Reactive Range Low Range: 0.00 to 1.60 mg/L (as P)
High Range: 0.0 to 32.6 mg/L (as P)
Phosphorous, Reactive Resolution 0.01 mg/L; 0.1 mg/L
Phosphorous, Reactive Accuracy Low Range: Ā±0.05 mg/L or Ā±4% of reading
High Range: Ā±0.5 mg/L or Ā±4% of reading
Phosphorous, Reactive Method Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th edition and EPA method 365.2 (low range)
Low Range: ascorbic acid method
High Range: vanadomolybdophosphoric acid method
Phosphorous, Total Range Low Range: 0.00 to 1.15 mg/L (as P)
High Range: 0.0 to 32.6 mg/L (as P)
Phosphorous, Total Accuracy Low Range: Ā±0.05 mg/L or Ā±6% of reading
High Range: Ā±0.5 mg/L or Ā±5% of reading
Phosphorous, Total Resolution Low Range: 0.01 mg/L; 0.1 mg/L
Phosphorous, Total Method Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th edition and EPA method 365.2 (low range)
Low Range: ascorbic acid method
High Range: vanadomolybdophosphoric acid method
Input Channels 1 pH electrode input and 5 photometer wavelengths
pH Electrode digital pH electrode (not included)
Photometer/Colorimeter Light Source 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters
Photometer/Colorimeter Light Detector silicon photodetector
Bandpass Filter Bandwidth 8 nm
Bandpass Filter Wavelength Accuracy Ā±1 nm
Cuvette Type round, 24.6 mm
Number of Methods 128 max.
GLP calibration data for connected pH electrode
Display 128 x 64 pixel LCD with backlight
Logging Type log on demand with user name and sample ID optional input
Logging Memory 1000 readings
Connectivity USB-A host for flash drive; micro-USB-B for power and computer connectivity
Power Supply 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Battery Type/Life 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Environment 0 to 50.0 ℃ (32 to 122.0 °F); 0 to 95% RH, non-condensing
Weight 1.0 kg (2.2 lbs.)
Dimensions 206 x 177 x 97 mm (8.1 x 7.0 x 3.8 in.)

Ordering Info

HI83314 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.

Manuals

Wastewater Multiparameter (with COD) Benchtop Photometer and pH meter – HI83314

Bestsellers

Digital Glass Body pH Electrode for General Purpose - HI11310

601 AED
The HI11310 is a glass body refillable double junction pH electrode with a built-in temperature sensor for temperature compensated measurements in a single probe design.Ā This probeĀ connects to the meter with a 3.5 mm connector. The 3.5 mm connector is proprietary to meters that use digital electrodes. The HI11310 electrode features a built-in microchip that stores sensor type, serial number, and calibration information.Ā It also features a single ceramic junction and the spherical pH sensing portion is made with high temperature glass. This design consideration is ideal for general purpose laboratory use.
  • Built-In Microchip
  • Spherical Glass Tip
  • Double Junction Reference
  • 3.5 mm Connector
Inquire Now

HALOĀ® BluetoothĀ® Wireless pH Meter for Flat Surfaces - HI14142

932 AED
The HI14142 HALO is an innovative pH electrode with BluetoothĀ® Smart technology designed for flat surfaces Our HALOĀ® Wireless pH Meter for Flat Surfaces has a flat tipped glass sensor and an open junction design to measure pH on flat surfaces or small volume samples. All readings are transmitted directly to your Apple or Android device or the edgeĀ®blu. One Press Connect Easily connect to the Hanna Lab App at the press of a button via Bluetooth wireless technology. Status Indicator Visible from a distance, the LED ā€œhaloā€ light flashes to indicate the status of the probe. Easy-to-Replace battery The HALOs CR2032 lithium ion battery is easily replaceable and lasts for approximately 500 hours.

Refillable Combination pH Electrode with BNC Connector - HI1131B

543 AED
The HI1131B is a glass body, refillable, double junction pH electrode with a BNC connector. This electrode has a single ceramic junction in the outer reference cell and the spherical pH sensing portion is made with high temperature glass. This design consideration is ideal for laboratory samples, beer and other liquid samples, as well as general purpose use.
  • Refillable Electrode
  • Double Junction Design
  • High Temperature (HT) Glass
Inquire Now

Foodcare pH Electrode for Meat Products with Quick Connect DIN Connector - FC2323

932 AED
The FC2323 is an amplified single junction pH electrode that has a built-in temperature sensor and connects to the meter with a quick connect DIN connector. This type of connector is proprietary to specific meters such as the HI98190. The FC2323 electrode featuresĀ a food grade plastic PVDF body, an open junction design with viscolene gel electrolyte, a sensing bulb made of low temperature glass, and a conical tip shape.Ā This design consideration, along with the optional stainless steel blades, is ideal for pH measurements in meat and semi-frozen food products. The recommended operating temperature range is fromĀ 0 to 50ā„ƒ.
  • Low Temperature Glass
  • Stainless Steel Piercing Blade (Optional)
  • Open Junction
Inquire Now

ChecktempĀ® 4 Folding Thermometer - HI151-Blue-No

121 AED
The HI151 ChecktempĀ® 4 is a folding thermistor thermometer that can measure temperature up to 300 ā„ƒ (572 ā„‰). These thermometers offer many advanced features including CAL-Check for an internal verification of the electronics and a motion sensor which eliminates the need of closing and reopening the probe when the meter enters a power saving mode. The Checktemp 4 is available in six different colors which can be assigned to different products to avoid cross contamination in the restaurant, kitchen or production facility.
  • CAL Check to verify meter performance
  • Waterproof to IP67 Standards
  • Wake on motion
Inquire Now

Spare Electrode for PICCOLOĀ® plus pH Tester - HI1295

277 AED
The HI1295 is a spare electrode for the HI98113 PICCOLOĀ® plus pH Tester.

Titanium Body pH Electrode for Wastewater with Quick Connect DIN Connector - HI12963

761 AED
The HI12963 is a titanium body, gel filled, single reference junction pH electrode that has a built-in temperature sensor and a quick connect DIN connector. This type of connector is proprietary to specific meters such as the HI98190 and HI991001. The HI12963 electrode is made with general purpose glass and has an extendable cloth type junction. The design considerations are ideal for measuring the pH in industrial, municipal, and environmental applications.
  • Titanium Body
  • Built-In Temperature Sensor
  • Quick Connect DINĀ 